Fat-water separated T1 mapping with inversion-prepared multi-echo MRI
Fortin, Marc-Antoine · container not stated · 2022
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
· OpenAlex · dissertation · 2022
Abstract
Knowing the oxygen level inside cancer tumors before radiotherapy can significantly affect the choice of treatment Currently, measuring oxygen levels in vivo before radiotherapy is not performed due to the lack of a simple and validated method that could be easily incorporated clinically. Recent work in Magnetic Resonance Imaging (MRI) reported that variations in fat T1 values could be used as an indirect and non-invasive method to measure tissue oxygenation.
For this purpose, the Magnetization Prepared Two Rapid Acquisition of Gradient Echoes (MP2RAGE) sequence designed for brain T1 mapping and 3-point Dixon to perform fat-water separation were adapted and combined to design a novel fat-water separated T1 mapping technique from a multi echo (ME) MP2RAGE sequence.Extensive assessment and validation of the accuracy and precision of the newly designed ME-MP2RAGE protocol were performed throughout this project.
First, from a qualitative assessment of the acquisition parameters, the original MP2RAGE protocol developed for T1 values between 1000 ms and 2000 ms (typical of brain tissue at 3 T) was modified to be optimized for a T1 range between 200 ms and 800 ms. A quantitative assessment of the theoretical accuracy and precision for the new short T1 MP2RAGE protocol was performed with numerical simulations to compare with the original brain T1 MP2RAGE protocol. Afterwards, the short T1 MP2RAGE protocol was experimentally validated against the original brain T1 MP2RAGE protocol and a reference T1 mapping technique in a homemade phantom with water and gadolinium-based contrast agent.
The experimental feasibility of a ME-MP2RAGE protocol for fat-water separation with 3-point Dixon was also verified. The fat-water separated T1 mapping technique from the short T1 ME-MP2RAGE protocol was then experimentally assessed with a second experiment and a homemade phantom with fat and water.
Finally, an extensive quantitative assessment of the fat-water separated T1 values calculated from the ME-MP2RAGE protocol and 3-point Dixon across a wide range of T1 values was performed inside another homemade phantoms including different concentrations of gadolinium-based contrast agent and fat fractions.The qualitative and quantitative assessments of the new short T1 MP2RAGE protocol showed higher theoretical accuracy and precision than the original brain T1 MP2RAGE protocol for the T1 range of interest.
Moreover, the T1 values calculated from the short T1 MP2RAGE showed a remarkably high correlation with the reference T1 mapping technique whereas the feasibility of the ME-MP2RAGE protocol was also demonstrated experimentally. The second experiment demonstrated that the ME-MP2RAGE protocol with 3-point Dixon was suitable for precise and uniform fat-water separated T1 mapping.
With the final experiment, the fat T1 values calculated from the ME-MP2RAGE protocol did not show any trend across different fat fractions and concentrations of gadolinium-based contrast agent whereas the water T1 values showed an expected decrease with increasing concentrations of gadolinium-based contrast agent and smaller but noticeable decrease with increasing fat fractions.In summary, the fat-water separated ME-MP2RAGE T1 mapping technique presented in this thesis can easily and rapidly produce accurate and precise fat-water separated T1 maps.
Moreover, the precision level measured in fat T1 values would make the technique suitable to measure variations in fat T1 values due to the presence of oxygen. Once validated, this new technique could be used as a non-invasive technique to measure oxygen level in cancer tumors