Magnetic field-induced variation in thermo luminscent dosimeter (tld-100) response during mri-guided radiotherrapy :P hantom study
Abhimanu Kumar, Harikumar Pallathadka · African Journal of Biomedical Research · 2026
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
African Journal of Biomedical Research · African Journals OnLine · article · 2026
Abstract
Background: The integration of magnetic resonance imaging (MRI) with linear accelerators (MR-Linacs) provides real-time tumor tracking but introduces complex magnetic fields that may alter the dose-response characteristics of standard dosimeters. Objective: To quantify the impact of a 1.5 T transverse magnetic field on the sensitivity and dose-response linearity of LiF:Mg,Ti (TLD-100) dosimeters. Methods: Sixty TLD-100 chips were irradiated using a 6 MV photon beam in a water-equivalent phantom. The dosimeters were divided into two groups: Group A (control, 0 T) and Group B (experimental, 1.5 T magnetic field applied perpendicular to the beam axis).
Dosimeters were exposed to varying dose levels (50 cGy to 500 cGy). Readouts were processed using a standard TLD reader, and dose-response curves were analyzed using linear regression. Results: Both groups exhibited excellent dose-response linearity (R2 > 0.998). However, Group B demonstrated a statistically significant over-response. At 200 cGy, the mean relative thermos luminescent (TL) yield for Group B was 3.4% higher than Group A (p < 0.05). The magnetic field effect factor (kB) was calculated as 1.034 ± 0.012. Conclusion: The presence of a 1.5 T magnetic field induces a measurable increase in the TL response of TLD-100 dosimeters.
This perturbation must be accounted for by applying appropriate correction factors during routine clinical dosimetry in MRgRT environments.