0_All_tabs.zip
Christina Maillet, M. Klein, Florence, Florian Le Bras, Frédéric Velard, Christine Guillaume, Sophie C. Gangloff, Gelle, Marie-Paule · Recherche Data Gouv France · 2023
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
Recherche Data Gouv France · OpenAlex · dataset · 2023
Abstract
An innovative process based on an oxygen non-thermal plasma (O2-NTP) generated in a sealed bag could be used for the sterilization of fragile medical devices. The purpose of this data collection was to assess the cytocompatibility of titanium (TA6V) and poly(etheretherketone) (PEEK) surfaces after O2-NTP sterilization compared to gamma-ray irradiation (GI) and autoclave (AC). This data collection contains the results of three hallmarks of cytocompatibility assessed on MG-63 osteoblast-like cells seeded on TA6V and PEEK disks sterilized by AC, GI and O2-NTP : cell's viability by WST1 assay, cell's proliferation by DNA quantification, and cytotoxicity by LDH assay.
Open the resolved locator · 10.57745/eggtwb
Metadata licence: OpenAlex metadata (CC0)