Electron Holography of Magnetic Materials

Takeshi Kasama, E. Rafal, Marco Beleggi · InTech eBooks · 2011

  • DOI 10.5772/22366
  • Publisher OpenAlex
  • Type book-chapter
  • Issued 2011

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

InTech eBooks · OpenAlex · book-chapter · 2011

Abstract

IntroductionTransmission electron microscopy (TEM) involves the use of high-energy (60-3000 keV) electrons that have passed through a thin specimen to record images, diffraction patterns or spectroscopic information from a region of interest.Many different TEM techniques have been developed over the years into highly sophisticated methodologies that have found widespread application across scientific disciplines.Because the TEM has an unparalleled ability to provide structural and chemical information over a range of length scales down to atomic dimensions, it has developed into an indispensable tool for scientists who are interested in understanding the properties of nanostructured materials and in manipulating their behavior (Smith, 2007).State-of-the-art TEMs are now equipped with spherical and chromatic aberration correctors and can provide interpretable image resolutions of 0.05 nm (Erni et al., 2009).However, in addition to conventional TEM techniques that can be used to provide structural and compositional information about materials, the TEM also allows magnetic and electrostatic fields in specimens to be imaged with nanometer spatial resolution.One of the most powerful techniques for providing this information is electron holography, which was originally proposed as a means to compensate for lens aberrations and to improve electron microscope resolution (Gabor, 1949).Electron holography is still the only technique that provides direct access to the phase shift of the electron wave that has passed through a thin specimen, in contrast to more conventional TEM techniques that record only spatial distributions of image intensity.Electron holography has only recently become widely available on commercial electron microscopes.The earliest studies using electron holography were restricted by the limited brightness and coherence of the tungsten filaments that were used as electron sources (Haine & Mulvey, 1952).The availability of high brightness, stable, coherent field emission electron guns now allows electron holography to be applied to a wide variety of materials such as quantum well structures, magnetic thin films, semiconductor devices, natural rocks and biominerals. www.intechopen.com Holography -Different Fields of Application 54Although there are several forms of electron holography, including in-line and scanning TEM-based electron holography (Cowley, 1992), the most commonly-used form is the TEM mode of off-axis electron holography, which involves using an electrostatic biprism that is inserted into the microscope column perpendicular to the electron beam (Möllenstedt & Düker, 1956).As described in detail below, the application of a voltage to the biprism results in overlap of the electron wave to create an interference fringe pattern.Other forms of electron holography are described elsewhere (e.g., Cowley & Spence, 1998).Off-axis electron holography can be divided further into two modes: (i) High-resolution electron holography, in which the interpretable resolution in a high-resolution TEM image can be improved by the use of phase plates to correct for microscope lens aberrations (Lichte, 1991;Lehmann et al., 1999) and (ii) Medium-resolution electron holography, in which magnetic and electrostatic fields in materials can be studied quantitatively, usually with nanometer spatial resolution (Tonomura, 1992;Dunin-Borkowski et al., 2004).Some of the most successful early examples of the application of medium-resolution electron holography include the experimental confirmation of magnetic flux quantization in superconducting toroids (i.e., the Aharonov-Bohm effect) and the study of magnetic flux vortices in superconductors (Tonomura et al., 1986;Matsuda et al, 1989).Here we concentrate on more recent applications of medium-resolution electron holography for the study of magnetic materials.This chapter begins with an outline of the basis and requirements for electron holography and the procedures that are typically used to obtain amplitude and phase information from electron holograms, including microscope calibrations for magnetic studies and the practical steps that are required to extract the magnetic contribution to the signal from a reconstructed phase image.We then provide selected examples of the application of electron holography to the characterization of magnetic domain structures in nanostructured materials.We describe a novel approach for measuring the magnetic moment of a chosen nanocrystal quantitatively, before discussing future opportunities for the development and application of electron holography for the study of magnetic materials and devices.

Basis of off-axis electron holographyOff-axis electron holography in the TEM involves the examination of a thin specimen using a highly coherent field emission electron source.A schematic diagram showing the t y p i c a l e l e c t r o n -o p t i c a l c o n f i g u r a t i o n i s s h o w n i n F i g .

1 a .T h e r e g i o n o f i n t e r e s t i s positioned so that it covers approximately half the field of view.A biprism (a thin conducting wire), which is usually located close to the first image plane, has a positive voltage applied to it in order to overlap a "reference" electron wave that has passed through vacuum with the electron wave that has passed through the specimen.The overlap region contains interference fringes and an image of the specimen, in addition to Fresnel fringes originating from the edge of the biprism wire (Figs 1b-e).The electronoptical configuration is equivalent to the use of two electron sources S 1 and S 2 , as shown in Fig.

1a (Dunin-Borkowski et al., 2004).By applying a larger voltage to the biprism wire, the separation of the two virtual sources and the width of the overlap region W increase (Figs 1b-d), placing a greater constraint on the spatial coherence of the source required to maintain sufficient interference fringe contrast.

Open the resolved locator · 10.5772/22366

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 04:11 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://www.intechopen.com/chapters/19623
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.