Effect of chelating agents on preconcentration of trace heavy metals in aqueous solutions by thermoresponsive polymer-mediated extraction
智一 田中, Tanaka, Tomokazu, 義三 三木, Yoshizo Miki · Institutional Repositories DataBase (IRDB) · 2012
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
Institutional Repositories DataBase (IRDB) · National Institute of Informatics · article · 2012
Abstract
In the preconcentration of trace heavy metals in aqueous solutions with poly (N-isopropylacrylamide) (PNIPAAm), the effect of chelating agents on the enrichment of analytes was investigated. The use of 8-Quinolinol (Oxine) instead of ammonium 1-pyrrolidinedithiocardamate (APDC) enabled ca. 80-fold and 56-fold enrichment for Cr and Cu, respectively. The preconcentration factors improved compared to previously obtained values by using APDC (ca. 38 fold for Cr and ca. 44 fold for Cu). This enhancement seems to be due to the decrease in the loss of metal chelates during the measurement of graphite-furnace atomic absorption spectrometry.
Since the loss, however, cannot be completely suppressed even by using Oxine, further investigation such as the use of different chelating agents is required for practical applications of the proposed method.