go-v.orgdoi

Seven Years of Left Ventricle Pacing Due to Mal-positioning Through Patent Foramen Ovale .

— · The Internet Journal of Cardiology · 2010

  • DOI 10.5580/82e
  • Publisher OpenAlex
  • Type article
  • Issued 2010

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

The Internet Journal of Cardiology · OpenAlex · article · 2010

Abstract

Pacing in the right ventricular apex is typically seen on the surface ECG as a Left Bundle Branch Block (LBBB) pattern, hence presence of a Right Bundle Branch Block (RBBB) pattern should raise a suspicion for atypical conduction pattern or lead malposition. Unintentional mal-positioning of the pacemaker lead in different locations , including left ventricle has been reported . This may be due to unintentional inter-atrial septal puncture or through a congenital defect such as patent foramen ovale (PFO), atrial or ventricular septal defects. Cannulation of the sub-clavian artery is another possibility. The exact incidence of a lead malposition is not known but is rare.

Patients may remain asymptomatic with normal lead function but are at an increased risk for thromboembolic events. We report a case of lead mal-positioning in the left ventricle through a PFO which was incidentally discovered years later.

Open the resolved locator · 10.5580/82e

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 01:04 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://doi.org/10.5580/82e
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.