go-v.orgdoi

Surgical Fire During Organ Procurement

— · The Internet Journal of Anesthesiology · 2009

  • DOI 10.5580/199c
  • Publisher Internet Scientific Publications
  • Type article
  • Issued 2009

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

The Internet Journal of Anesthesiology · Internet Scientific Publications · article · 2009

Abstract

Fire remains a dire complication in anesthesiology despite numerous advisories and education. Here, a case of fire occurring during organ procurement is described. Root cause analysis revealed that lack of communication and using an electrocautery in the presence of an alcohol-moisten sponges triggered fire. It resulted in a total loss of breathing circuit and 2 degree burns to the organ donor. The origin of fire is discussed with special emphasis on the role of communication between operating team members.

Open the resolved locator · 10.5580/199c

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 16:22 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://doi.org/10.5580/199c
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.