Data associated with 'The Influence of Kinematic Conditions and Variations in Component Positioning on the Severity of Edge Loading andWear of Ceramic-on-Ceramic Hip Bearings'
Murat Ali, Mazen Al‐Hajjar, Fisher, John, Louise M. Jennings · University of Leeds · 2020
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
University of Leeds · University of Leeds · dataset · 2020
Abstract
Data associated with the paper "The Influence of Kinematic Conditions and Variations in Component Positioning on the Severity of Edge Loading and Wear of Ceramic-on-Ceramic Hip Bearings". This dataset contains: - Figure 4. Dynamic separation (mean +/- 95% confidence limits, n = 3) of 36 mm diameter BIOLOX®delta ceramic-on-ceramic bearings under two axes and three axes of rotation conditions with 45 degree and 65 degree cup inclination angle conditions and 1, 2, 3 and 4 mm of medial-lateral translational mismatch conditions. - Figure 5.
Load at the rim at separation (mean +/- 95% confidence limits, n = 3)of 36 mm diameter BIOLOX®delta ceramic-on-ceramic bearings under two axes and three axes of rotation conditions at 45 degree and 65 degree cup inclination angles with 1, 2, 3 and 4 mm of medial-lateral translational mismatch. - Figure 6. Severity of edge loading (mean ± 95% confidence limits, n = 3) of 36 mm diameter BIOLOX®delta ceramic-on-ceramic bearings under two axes and three axes of rotation conditions with 45 degree and 65 degree cup inclination angles and 1, 2, 3 and 4 mm of medial-lateral translational mismatch conditions (offset deficiency). - Figure 7.
Wear rates (mean ± 95% confidence limits, n = 6) of 36 mm diameter BIOLOX®delta ceramicon-ceramic bearings under two axes and three axes of rotation conditions with 45 degree and 65 degree cup inclination angles and 4 mm of medial-lateral translational mismatch conditions.
Open the resolved locator · 10.5518/616