Gas Exchange Efficiency of an Oxygenator with Integrated Pulsatile Displacement Blood Pump for Neonatal Patients

Peter Schlanstein, Ralf Borchardt, Ilona Mager, Thomas Schmitz‐Rode, Ulrich Steinseifer, Jutta Arens · The International Journal of Artificial Organs · 2014

  • DOI 10.5301/ijao.5000293
  • Publisher SAGE Publishing
  • Type article
  • Issued 2014

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

The International Journal of Artificial Organs · SAGE Publishing · article · 2014

Abstract

Oxygenators have been used in neonatal extracorporeal membrane oxygenation (ECMO) since the 1970s. The need to develop a more effective oxygenator for this patient cohort exists due to their size and blood volume limitations. This study sought to validate the next design iteration of a novel oxygenator for neonatal ECMO with an integrated pulsatile displacement pump, thereby superseding an additional blood pump. Pulsating blood flow within the oxygenator is generated by synchronized active air flow expansion and contraction of integrated silicone pump tubes and hose pinching valves located at the oxygenator inlet and outlet.

The current redesign improved upon previous prototypes by optimizing silicone pump tube distribution within the oxygenator fiber bundle; introduction of an oval shaped inner fiber bundle core, and housing; and a higher fiber packing density, all of which in combination reduced the priming volume by about 50% (50 to 27 mL and 41 to 20 mL, respectively). Gas exchange efficiency was tested for two new oxygenators manufactured with different fiber materials: one with coating and one with smaller pore size, both capable of long-term use (OXYPLUS® and CELGARD®). Results demonstrated that the oxygen transfer for both oxygenators was 5.3-24.7 mlO2/min for blood flow ranges of 100-500 mlblood/min.

Carbon dioxide transfer for both oxygenators was 3.7-26.3 mlCO2/min for the same blood flow range. These preliminary results validated the oxygenator redesign by demonstrating an increase in packing density and thus in gas transfer, an increase in pumping capacity and a reduction in priming volume.

Open the resolved locator · 10.5301/ijao.5000293

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 12:19 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://doi.org/10.5301/ijao.5000293
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.