go-v.orgdoi

Corrosion study of metallic biomaterials in simulated body fluid

Hamid Reza Asgari Bidhendi, M. Pouranvari · Metalurgija-Journal of Metallurgy · 2011

  • DOI 10.30544/384
  • Publisher Association of Metallurgical Engineers of Serbia
  • Type article
  • Issued 2011

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

Metalurgija-Journal of Metallurgy · Association of Metallurgical Engineers of Serbia · article · 2011

Abstract

Titanium alloys and stainless steel 316L are still the most widely used biomaterials for implants despite emerging new materials for this application. There is still someambiguity in corrosion behavior of metals in simulated body fluid (SBF). This paper aims at investigating the corrosion behavior of commercially pure titanium (CP-Ti), Ti–6Al–4V and 316LVM stainless steel (316LVM) in SBF (Hank’s solution) at37 ºC using the cyclic polarization test. Corrosion behavior was described in terms of breakdown potential, the potential and rate ofcorrosion, localized corrosion resistance, andbreakdown repassivation.

The effects of anodizing on CP-Ti samples and the passivation on the 316LVM were studied in detail. It was shown that CP-Ti exhibited superior corrosion properties compared to Ti–6Al–4V and 316LVM.

Open the resolved locator · 10.30544/384

Metadata licence: OpenAlex metadata (CC0)