Concurrent Validity of Isokinetic Hip and Knee Strength Measurements Using the UGO Exoskeleton in Healthy Young Adults: Cross-Sectional Validation Study

Hao Xie, Minghui Ding, Peiming Chen, Shamay S. M. Ng, Chuhuai Wang, Jiang-Li Zhao · JMIR Rehabilitation and Assistive Technologies · 2026

  • DOI 10.2196/99564
  • Publisher JMIR Publications
  • Type article
  • Issued 2026

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

JMIR Rehabilitation and Assistive Technologies · JMIR Publications · article · 2026

Abstract

BACKGROUND: The UGO lower limb exoskeleton rehabilitation robot is designed for patients with lower limb motor dysfunction, and it can be used to assess the peak torque of hip and knee flexors and extensors. OBJECTIVE: The primary aim of this study was to establish the concurrent validity of peak isokinetic torque using the UGO exoskeleton and Humac Norm dynamometers in healthy young adults. METHODS: Twenty healthy young adults were enrolled in this study. All participants underwent isokinetic muscle strength tests using both the UGO exoskeleton and the Humac Norm dynamometers. Peak torque (Nm) of the hip and knee flexors and extensors was measured.

The Pearson correlation coefficient was used to determine concurrent validity. Bland-Altman analysis was used to assess the agreement between the 2 devices.

RESULTS: At the knee, peak torque of the extensors measured by the UGO exoskeleton correlated strongly with that measured by the Humac Norm dynamometer (right: r=0.662; P=.001; left: r=0.768; P<.001), with 95% limits of agreement (LoA) ranging from -22.89 to 82.21 Nm and from -24.60 to 75.65 Nm, respectively; the right knee flexors showed a moderate correlation (r=0.547; P=.01), with 95% LoA ranging from -26.27 to 36.01 Nm, whereas the left knee flexors showed no significant association (r=0.347; P=.12), with 95% LoA ranging from -2.40 to 75.02 Nm.

At the hip, peak torque of the flexors correlated strongly on the right (r=0.615; P=.003) and moderately on the left (r=0.443; P=.04), with 95% LoA ranging from -15.52 to 86.57 Nm and from -12.53 to 98.50 Nm, respectively, whereas the bilateral hip extensors showed no significant association (right: r=0.337; P=.14; left: r=0.391; P=.20), with 95% LoA ranging from -51.57 to 76.80 Nm and from -28.19 to 79.59 Nm, respectively. CONCLUSIONS: In healthy young adults, the UGO exoskeleton demonstrated moderate to strong concurrent validity for peak torque of the bilateral knee extensors and hip flexors, but not for the nondominant knee flexors or bilateral hip extensors.

The device may serve as a convenient torque assessment tool for selected muscle actions, but systematic overestimation and posture-related differences must be considered, and validation in the intended clinical population of patients with neurological conditions is required in further studies.

Open the resolved locator · 10.2196/99564

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 12:22 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://rehab.jmir.org/2026/1/e99564
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.