The Accuracy of Wrist-Worn Photoplethysmogram–Measured Heart and Respiratory Rates in Abdominal Surgery Patients: Observational Prospective Clinical Validation Study

Jonna A. van der Stam, Eveline H. J. Mestrom, Jai Scheerhoorn, Fleur E N B Jacobs, Simon Willem Nienhuijs, Arjen‐Kars Boer, Natal A. W. van Riel, Helma M. de Morree, A. Bonomi, Volkher Scharnhorst, R. Arthur Bouwman · JMIR Perioperative Medicine · 2023

  • DOI 10.2196/40474
  • Publisher JMIR Publications
  • Type article
  • Issued 2023

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

JMIR Perioperative Medicine · JMIR Publications · article · 2023

Abstract

BACKGROUND: Postoperative deterioration is often preceded by abnormal vital parameters. Therefore, vital parameters of postoperative patients are routinely measured by nursing staff. Wrist-worn sensors could potentially provide an alternative tool for the measurement of vital parameters in low-acuity settings. These devices would allow more frequent or even continuous measurements of vital parameters without relying on time-consuming manual measurements, provided their accuracy in this clinical population is established.

OBJECTIVE: This study aimed to assess the accuracy of heart rate (HR) and respiratory rate (RR) measures obtained via a wearable photoplethysmography (PPG) wristband in a cohort of postoperative patients. METHODS: ). The wearable obtained HR and RR measurements were compared to those of the reference monitor in the postanesthesia or intensive care unit. Bland-Altman and Clarke error grid analyses were performed to determine agreement and clinical accuracy. RESULTS: Data were collected for a median of 1.2 hours per patient.

With a coverage of 94% for HR and 34% for RR, the device was able to provide accurate measurements for the large majority of the measurements as 98% and 93% of the measurements were within 5 bpm or 3 rpm of the reference signal. Additionally, 100% of the HR and 98% of the RR measurements were clinically acceptable on Clarke error grid analysis. CONCLUSIONS: The wrist-worn PPG device is able to provide measurements of HR and RR that can be seen as sufficiently accurate for clinical applications. Considering the coverage, the device was able to continuously monitor HR and report RR when measurements of sufficient quality were obtained.

TRIAL REGISTRATION: ClinicalTrials.gov NCT03923127; https://www.clinicaltrials.gov/ct2/show/NCT03923127.

Open the resolved locator · 10.2196/40474

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 14:48 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://periop.jmir.org/2023/1/e40474
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.