DIFFERENTIAL COLORIMETRY FOR NUCLEAR PROCESS STREAMS.
Dow Chemical Co., Denver, Colo. Rocky Flats Div., J. Martinez, Geoff Hart · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · 1968
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · Office of Scientific and Technical Information · report · 1968
Abstract
An in-line process differential colorimeter, with automatic check system, has been developed with thhe capability of analyzing Pu(IV) in the range of 0.1 to 2.5 grams per liter. The instrument measures the difference in light transmittance at two wavelengths, 580 and 660 nanometers. These wavelengths are a characteristics of Pu(IV), ad the difference in transmittance is a measure of the plutonium concentration.