Materials survey for a radioisotope heat source for an artificial heart device
S.E. Bramer, USDOE · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · 1971
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · Office of Scientific and Technical Information · report · 1971
Abstract
venting of the helium gas generated by fuel decay and also to allow decay of the generated radon-220 gas to allowable levels. Pressure relief device tests verified its flow stability and fuel fines retention capability. Radon attenuation tests indicated a 0.001-in. I.D. capillary is required to reduce the radon levels to acceptable levels if it is assumed that all generated radon is released from the fuel.
Safety design verification tests, including granite impact, bullet impact, crushing, fire, and postaccident oxidation, indicated both the baseline and Hafnalloy designs to survive all accident environments except one by containing the fuel simulant in at least one totally integral capsule component. Neither design was capable of containing the fuel simulant under a 2000 ft-lb 30-06 rifle bullet impact. Further studies in this area as related to heat source design are required. (auth)
Open the resolved locator · 10.2172/4480997