A Mass-Spectrometer-Tube-Unit Prealigning and Positioning Jig for the Consolidated 21-102 Spectrometer
Louis P. Cecchini, James George · OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · 1952
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · Office of Scientific and Technical Information · report · 1952
Abstract
The construction of a jig for the external prealignment and positioning of mass spectrometer tube units (MSTU) in the Consolidated 21-102 mass spec trometer is described.Functions and illustrations of the constituent parts are given.The procedures and precautions for installing and aligning the MSTU in the jig are enumerated as are the procedures and precautions for in serting the MSTU-jig assembly into the magnet gap.By employing the jig de scribed above, the MSTU is aligned more accurately and in less time than is possible with the conventional procedures.Strains inherently created at definite points when the recommended procedures are followed, are minimized as a result of the increased support provided by the jig for the MSTU.It is to be expected that the more nearly optimum alignment of the MSTU, easily achieved with the use of the jig, will insure the attainment of the desired resolution, sensitivity, and reproducibility of ion patterns.
Open the resolved locator · 10.2172/4391981
Metadata licence: OpenAlex metadata (CC0)