女子大生を対象とした血圧測定におけるカフ幅の検討

Atsuko MASUDA, Ayako Ninomiya, 仁子 小泉 · Institutional Repositories DataBase (IRDB) · 2007

  • DOI 10.18933/00000018
  • Publisher National Institute of Informatics
  • Type article
  • Issued 2007

Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.

Institutional Repositories DataBase (IRDB) · National Institute of Informatics · article · 2007

Abstract

The blood pressure cuff must have the correct width for the arm circumference, for if it is too narrow, the blood pressure reading will be erroneously higher ; if it is too wide, the reading may be erroneously lower. The purpose of this study is to identify the proper cuff width for measuring blood pressure of young adult women precisely. Using the cuff of 12cm (for adults) as the standard size and 9cm (for children) in width, we measured the blood pressure of 41 female university students. The mean arm circumference was 23.6cm requiring a cuff width of 9.4cm, because the cuff width should be 40% of the arm circumference.

Therefore, we found that the standard cuff width of 12cm was too large for the subjects of our study. And, with standard cuff width (12cm), the mean systolic and diastolic blood pressure were underestimated by 7.7mmHg and 9.8mmHg, respectively. Such underestimation may lead to a misdiagnosis of pre-eclampsia, particularly in lean pregnant women.

Open the resolved locator · 10.18933/00000018

Metadata licence: OpenAlex metadata (CC0)

Registry
OpenAlex DOI index + API
Fetched
2026-11-20 04:13 UTC
Cache state
not cached yet (bulk-index+openalex)
Locator
https://ryotokuji-u.repo.nii.ac.jp/records/26
Metadata licence
OpenAlex metadata (CC0)
Scope
Metadata as returned by the resolver; abstracts are never generated here.