Wavelet analysis of active cavitation detector output
Kofi A. Anguah, E. Carr Everbach · The Journal of the Acoustical Society of America · 2006
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
The Journal of the Acoustical Society of America · Acoustical Society of America · conference-abstract · 2006
Abstract
Active cavitation detectors (ACDs) have provided useful information concerning the dynamics of microbubbles exposed to intense pulsed ultrasound. In an examination of the effects of nearby solid surfaces on the asymmetrical collapse of bubbles, ACD signals consisting of 60-cycle 30-MHz tone bursts of <0.1 MPa peak negative pressure returned complex echoes for individual bubbles and groups of bubbles. The Gauss-Harmite wavelet transform was used to analyze these signals to differentiate between symmetrical and asymmetrical collapse signatures, with the goal of identifying the percentage of asymmetrical bubble collapses in a cavitation field. [Work supported by NIH R-15 EB004630-01.]
Open the resolved locator · 10.1121/1.4786355
Metadata licence: OpenAlex metadata (CC0)