Two-microphone measurement of the power-reflection coefficient in human ears
William J. Murphy, Arnold Tubis, Glenis R. Long · The Journal of the Acoustical Society of America · 1987
Not cached yet. This record was fetched live for this request and is stored after the lookup. Reopening the page should serve the cached copy.
The Journal of the Acoustical Society of America · Acoustical Society of America · article · 1987
Abstract
For each of eight normal hearing and eight abnormal hearing subjects, the acoustic intensity and power reflection coefficient spectra are measured using a two-microphone probe with probe tips ≈ 8 mm downstream from the point of sound delivery [W. J. Murphy, A. Tubis, and G. R. Long, J. Acoust. Soc. Am. Suppl. 1 81. S75 (1987)]. The tube of an Etymotic Research ER-2 sound delivery system passes through one port and the probe tips from a pair of ER-7 microphones pass through the second port of a latex ear mold. Results in the frequency range of 4–12 kHz are compared with previous measurements by Stinson et al. [J. Acoust. Soc. Am. 72, 766–773 (1982)], Hudde [J. Acoust. Soc. Am.
73, 242–247 (1983)], and Stinson [J. Acoust. Soc. Am. 77, 386–393 (1985); 79, 1003–1009 (1986)]. Hearing thresholds are also determined in order to identify any significant correlations in the objective and psychophysical measure. [Work supported by a grant from the Deafness Research Foundation.]
Open the resolved locator · 10.1121/1.2024948
Metadata licence: OpenAlex metadata (CC0)